Changeflow GovPing Government General Philips Spectral CT on Rails, Model 728334 - Cl...
Urgent Enforcement Amended Final

Philips Spectral CT on Rails, Model 728334 - Class II Recall (Z-1825-2026)

Favicon for changeflow.com FDA Device Recalls (Class II)
Filed
Detected
Email

Summary

Philips Medical Systems recalled 3 Spectral CT on Rails systems (Model 728334, Serial Numbers 1005–1007, Software Version 5.1.X) after identifying three software defects: gantry positioning failure during continuous CT scans, failure of a Select Previous Surview prompt to appear, and the Define Head Area option becoming grayed out under certain dose settings. Philips notified consignees on March 7, 2026 with specific workarounds for each issue and will dispatch Field Service Engineers to install a corrective software update (FCO72800830).

Published by Philips on accessdata.fda.gov . Detected, standardized, and enriched by GovPing. Review our methodology and editorial standards .

What changed

Philips Medical Systems initiated a Class II medical device recall for 3 Spectral CT on Rails systems (Model 728334) distributed in Minnesota, USA and in France and the Netherlands. The recall addresses three distinct software defects identified in Version 5.1.X: (1) potential gantry failure to advance after pressing Go during continuous CT scans, (2) a missing Select Previous Surview user prompt, and (3) graying out of the Define Head Area option when certain Brain Helical acquisition dose settings are active.

Facilities holding affected units (Serial Numbers 1005, 1006, 1007) must immediately inventory to identify in-scope systems, apply the specified workarounds from the Instructions for Use for each software issue, circulate the Urgent Medical Device Correction Letter to all device users, and retain the letter with each system until a Philips Field Service Engineer performs the corrective software installation (FCO72800830). Clinical engineering and radiology operations staff should coordinate on workaround protocols to maintain imaging availability during the interim period.

What to do next

  1. Philips notified consignees on about 03/07/2026 via letter. Consignees were instructed to identify any affected systems in inventory and specifically for each issue: 1. Issue #1: Ensure to follow the instructions mentioned in the IFU Section 4 Preparing for an Examination and Sub section Gantry Operation. 2. Issue #2: Avoid the Select Previous Surview Unavailable issue, manually remove the patient's middle name while importing the patient detail from HIS/RIS. 3. Issue #3: if Define Head Area option is grayed out with DoseRight Index (DRI) on, review the dose levels displayed on the User Interface, confirm and adjust the dose settings to appropriate levels before executing the scan as instructed in the IFU.
  2. Consignees were also instructed to complete and return the response form, to circulate the Urgent Medical Device Correction Letter to all users of the affected device so that they are aware of the issues, and retain the letter with affected system(s) until a solution is installed on each system; ensure the letter is in a place likely to be seen/viewed.
  3. Philips will contact each consignee to schedule a time for a Philips Field Service Engineer (FSE) to visit each site and install the solution to resolve the issues (reference FCO72800830).

Archived snapshot

Apr 20, 2026

GovPing captured this document from the original source. If the source has since changed or been removed, this is the text as it existed at that time.

Class 2 Device Recall Philips Spectral CT on Rails

- Databases

| |

| 510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards |

| CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | |

| New Search | ) |

| | Class 2 Device Recall Philips Spectral CT on Rails | |

Date Initiated by Firm March 07, 2026
Date Posted April 16, 2026
Recall Status Open , Classified
Recall Number Z-1825-2026
Recall Event ID 98588
510(K)Number K212875
Product Classification System, x-ray, tomography, computed - Product Code JAK
Product Philips Spectral CT on Rails. Model Number: 728334.
Code Information Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
FEI Number 3017726341
Recalling Firm/
Manufacturer PHILIPS MEDICAL SYSTEMS

222 Jacobs St
Cambridge MA 02141-2289 |
| For Additional Information Contact | Philips Customer Care Solutions Center
800-722-9377 |
| Manufacturer Reason
for Recall | Philips has identified three software issues:
1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go.
2. Potential where the message of Previous Surview Exists Select Previous Surview? or should display but does not appear.
3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out. |
| FDA Determined
Cause | Under Investigation by firm |
| Action | Philips Medical notified consignees on about 03/07/2026 via letter. Consignees were instructed to identify any affected systems in inventory and specifically for each issue:
1. Issue #1: Ensure to follow the instructions mentioned in the IFU Section 4 Preparing for an Examination and Sub section Gantry Operation.
2. Issue #2: Avoid the Select Previous Surview Unavailable issue, manually remove the
patient's middle name while importing the patient detail from HIS/RIS.
3. Issue #3: if Define Head Area option is grayed out with DoseRight Index (DRI) on, review
the dose levels displayed on the User Interface, confirm and adjust the dose settings to
appropriate levels before executing the scan as instructed in the IFU.

Consignees were also instructed to complete and return the response form, to circulate the Urgent Medical Device Correction Letter to all users of the affected device so that they are aware of the issues, and retain the letter with affected system(s) until a solution is installed
on each system; ensure the letter is in a place likely to be seen/viewed.

Philips will contact each consignee to schedule a time for a Philips Field Service Engineer (FSE) to visit each site and install the solution to resolve the issues (reference FCO72800830). |
| Quantity in Commerce | 3 systems |
| Distribution | Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands. |
| Total Product Life Cycle | TPLC Device Report |

| 1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes. | |

| 510(K) Database | 510(K)s with Product Code = JAK |

Get daily alerts for FDA Device Recalls (Class II)

Daily digest delivered to your inbox.

Free. Unsubscribe anytime.

About this page

What is GovPing?

Every important government, regulator, and court update from around the world. One place. Real-time. Free. Our mission

What's from the agency?

Source document text, dates, docket IDs, and authority are extracted directly from Philips.

What's AI-generated?

The summary, classification, recommended actions, deadlines, and penalty information are AI-generated from the original text and may contain errors. Always verify against the source document.

Last updated

Classification

Agency
Philips
Filed
March 7th, 2026
Instrument
Enforcement
Branch
Executive
Legal weight
Binding
Stage
Final
Change scope
Substantive
Document ID
Recall No. Z-1825-2026 / Recall Event ID 98588

Who this affects

Applies to
Medical device makers Healthcare providers
Industry sector
3345 Medical Device Manufacturing
Activity scope
Medical device recall Software defect response Field service correction
Geographic scope
United States US

Taxonomy

Primary area
Medical Devices
Operational domain
Clinical Operations
Compliance frameworks
GxP
Topics
Healthcare Product Safety

Get alerts for this source

We'll email you when FDA Device Recalls (Class II) publishes new changes.

Free. Unsubscribe anytime.

You're subscribed!