Health Canada has issued a Type II recall for Abbott Point of Care Inc.'s i-STAT EG7+ Cartridge due to a manufacturing defect causing inaccurate pH and pCO2 readings. Affected lots (N25209, Model 03P76-25) may report higher than expected pCO2 and lower than expected pH results, potentially leading to inappropriate clinical interventions. Healthcare providers should immediately discontinue use of these cartridges for pH and pCO2 testing and contact the manufacturer for additional information.

What changed

Health Canada has issued a Type II medical device recall for Abbott Point of Care Inc.'s i-STAT EG7+ Cartridge due to a manufacturing defect affecting specific lot N25209 (Model 03P76-25). The cartridges may report higher than expected pCO2 and lower than expected pH results, potentially presenting a pattern suggestive of severe respiratory acidosis.\n\nHealthcare providers using these cartridges should immediately discontinue their use for pH and pCO2 testing and contact Abbott Point of Care Inc. for replacement products. Failure to identify inaccurate results could lead to unnecessary or inappropriate clinical interventions based on misleading diagnostic data. This recall requires urgent attention from all healthcare facilities performing point-of-care blood gas analysis.

What to do next

  1. Discontinue use of affected i-STAT EG7+ Cartridges for pH and pCO2 testing
  2. Contact Abbott Point of Care Inc. for replacement products
  3. Report any health or safety concerns to Health Canada

Archived snapshot

Apr 16, 2026

GovPing captured this document from the original source. If the source has since changed or been removed, this is the text as it existed at that time.

Health product recall

i‑STAT EG7+ Cartridge

Brand(s)

Abbott Point Of Care Inc.

Last updated

2026-04-15

Summary

Product i‑STAT EG7+ Cartridge Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information.

Audience Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
i‑STAT EG7+ Cartridge N25209 03P76-25

Issue

Cartridges from the specific lots listed may report higher than expected pCO 2 and lower than expected pH results due to a manufacturing issue which may present a pattern suggestive of severe respiratory acidosis.  If such results are acted upon without consideration of the patient's overall clinical presentation, this could lead to unnecessary or inappropriate clinical interventions.  Use of the cartridges should be discontinued for pH and pCO 2.

Additional information

Details

Original published date:

2026-04-15

Alert / recall type Health product recall Category Health products - Medical devices - Chemistry Companies
| Abbott Point Of Care Inc. |
| 100 And 200 Abbott Park Rd., Abbott Park, Illinois, United States, 60064 |
Published by Health Canada Audience Healthcare Recall class Type II Recall date

2026-04-02

Identification number RA-81897

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe

Report a health or safety concern

Get daily alerts for Health Canada Recalls & Safety Alerts

Daily digest delivered to your inbox.

Free. Unsubscribe anytime.

About this page

What is GovPing?

Every important government, regulator, and court update from around the world. One place. Real-time. Free. Our mission

What's from the agency?

Source document text, dates, docket IDs, and authority are extracted directly from Health Canada.

What's AI-generated?

The summary, classification, recommended actions, deadlines, and penalty information are AI-generated from the original text and may contain errors. Always verify against the source document.

Last updated

Classification

Agency
Health Canada
Filed
April 2nd, 2026
Instrument
Enforcement
Legal weight
Binding
Stage
Final
Change scope
Substantive
Document ID
RA-81897

Who this affects

Applies to
Healthcare providers Medical device makers
Industry sector
3345 Medical Device Manufacturing
Activity scope
Point-of-care diagnostics Blood gas analysis Medical device recall
Geographic scope
Canada CA

Taxonomy

Primary area
Medical Devices
Operational domain
Clinical Operations
Compliance frameworks
GxP
Topics
Product Safety Healthcare

Get alerts for this source

We'll email you when Health Canada Recalls & Safety Alerts publishes new changes.

Free. Unsubscribe anytime.

You're subscribed!