Heart Pump Device Correction - Impella RP SmartAssist Sensor Drift

Favicon for www.fda.gov FDA Medical Device Recalls
Detected April 3rd, 2026
Email

Summary

FDA identified a Class I recall of Abiomed Impella RP with SmartAssist and Impella RP Flex with SmartAssist heart pumps due to differential pressure sensor drift causing inaccurate pump flow readings on the Automated Impella Controller. The correction involves updated Instructions for Use requiring clinicians to reference P-level flow rate tables instead of AIC display values. Affected product codes: 0046-0035 and 1000323.

What changed

Abiomed issued an Urgent Medical Device Correction for Impella RP heart pumps after identifying that the differential pressure (dP) sensor may malfunction and cause values to drift over time. This drift affects AIC display readings for pump flow, PA Placement Signal, PAPi, and CV Placement Signal. FDA classified this as the most serious recall type. The correction updates clinical instructions to use P-level flow rate ranges from the IFU tables rather than relying on the potentially inaccurate AIC display values.

Healthcare facilities using these devices must immediately distribute the correction notice to all relevant clinical staff, post it in visible areas, and contact any facilities that received forwarded products. Clinicians should monitor hemodynamics with approved diagnostic devices, verify Impella positioning with imaging, and compare AIC display values against the IFU P-level flow rate tables. Any abrupt changes in displayed flow or discrepancy from expected IFU values requires reassessment of pump positioning and patient conditions.

What to do next

  1. Distribute the Urgent Medical Device Correction letter to all clinical personnel managing, using, or transporting Impella RP devices
  2. Monitor patient hemodynamics with approved diagnostic devices and verify Impella positioning with imaging before clinical interventions
  3. Refer to P-level flow rate tables in updated IFUs and compare against AIC display values; flag any discrepancies for clinical review

Penalties

FDA identifies this as the most serious recall type; devices may cause serious injury or death if used without following updated instructions

Source document (simplified)


The affected products and recommendations for what to do with the devices below have not changed. This recall **involves updating instructions for using devices, and does not involve removing them from where they are used or sold.* The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without following the updated instructions.*

Affected Product

The FDA is aware that Abiomed has issued an Urgent Medical Device Correction to notify affected customers of updated use instructions for all Impella RP with SmartAssist and Impella RP Flex with SmartAssist devices.

Affected devices:

| Impella Product Code | Impella Product Description | Unique Device Identifier |
| --- | --- | --- |
| 0046-0035 | Impella RP with SmartAssist | 00813502011869 |
| 1000323 | Impella RP Flex with SmartAssist | 00813502012811 |

What to Do

Monitor patient hemodynamics with an approved diagnostic device and verify Impella positioning with imaging. Clinicians should refer to the P-level flow rates listed in the Instructions for Use (as shown below) and rely on these values rather than the AIC display values.

On January 27, Abiomed sent all affected customers a letter recommending clinicians take the following actions:

  • Continue to monitor patient hemodynamics with approved diagnostic devices and verify Impella positioning with imaging before clinical interventions.
  • Refer to the P-level flow rates listed in the Instructions for Use (IFUs), as shown in the tables below, and rely on these values rather than the AIC display.
    • The values in the IFU reflect the flow rate range of the pump on defined pressure gradients under controlled conditions.
    • Clinicians should be aware that actual pump flow depends on preload, afterload, and can vary due to suction, incorrect positioning, or presence of thrombus in the inlet.
  • Trend pump metrics to monitor pump performance and to identify sudden change, noting that abrupt changes in the displayed flow rate need to be assessed both against the flow rate listed in the IFU and across temporal trends. An abrupt change in flow or abrupt discrepancy from the expected flow rate listed in the IFU may indicate a need to reassess pump positioning and performance, patient conditions, and clinical hemodynamics.
  • Distribute this information to appropriate personnel at your facility who manage, transport, stock, store, or use the affected product.
  • If any of the affected products have been forwarded to another facility, contact that facility and provide them with this notice.
  • Post a copy of this notice in a visible area for awareness. P-Level Flow Rates Note: Flow rate depends on preload and afterload and can vary due to suction or incorrect positioning.

Impella RP Flex with SmartAssist

| P-Level | Flow Rate (L/min) |
| --- | --- |
| P-0 | 0.0 |
| P-1 | 0.0-0.9 |
| P-2 | 0.0-1.3 |
| P-3 | 0.0-1.8 |
| P-4 | 0.7-2.3 |
| P-5 | 1.3-2.6 |
| P-6 | 2.0-3.0 |
| P-7 | 2.7-3.4 |
| P-8 | 3.2-3.7 |
| P-9 | 3.7-4.2 |
Impella RP with SmartAssist

| P-Level | Flow Rate (L/min) |
| --- | --- |
| P-0 | 0.0 |
| P-1 | 0.0-1.2 |
| P-2 | 0.0-1.6 |
| P-3 | 0.0-2.0 |
| P-4 | 1.3-2.9 |
| P-5 | 1.6-3.1 |
| P-6 | 2.4-3.5 |
| P-7 | 3.0-4.0 |
| P-8 | 3.4-4.2 |
| P-9 | 3.9-4.4 |

Reason for Correction

Abiomed has identified that the differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift over time. If sensor drift occurs, the Automated Impella Controller (AIC) may display inaccurate information, including pump flow readings, Pulmonary Artery (PA) Placement Signal, Pulmonary Artery Pulsatility Index (PAPi), and Central Venous (CV) Placement Signal.

The dP sensor measures the pressure difference between the inlet and outlet of the pump and is used to calculate these displayed values. There is no way to confirm error in the flow calculation directly from the pump signals when sensor drift occurs. Inaccurate displayed information on the AIC has resulted in the user making clinical interventions including inaccurate P-level adjustments or unnecessary pump exchanges. However, the dP sensor does not affect the device’s ability to provide hemodynamic support.

As of January 15, Abiomed has reported 22 serious injuries and no deaths associated with this issue.

Device Use

The Impella RP with SmartAssist and Impella RP Flex with SmartAssist are temporary heart pumps intended to provide right heart support in patients with acute right heart failure. The device is used in hospital settings by trained health care professionals.

Contact Information

Customers in the U.S. with adverse reactions, quality problems, or questions about this issue should contact Abiomed, Inc. at onemed-field-actions@its.jnj.com.

Additional FDA Resources

Additional Company Resources

Company-provided information is posted here by the FDA as a public service.

Unique Device Identifier (UDI)

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.

Timeline of Communication Updates

| Date | Actions |
| --- | --- |
| 04/03/2026 | The FDA updated this communication to inform the public that this issue has been classified as a Class I Recall. |
| 02/03/2026 | The FDA issued an Early Alert communication to notify the public of a potentially high-risk device issue. |

  • ## Content current as of:

04/03/2026

  • Regulated Product(s)

    • Medical Devices

Named provisions

P-Level Flow Rates Impella RP Flex with SmartAssist Impella RP with SmartAssist

Source

Analysis generated by AI. Source diff and links are from the original.

Classification

Agency
FDA
Instrument
Enforcement
Legal weight
Binding
Stage
Final
Change scope
Substantive

Who this affects

Applies to
Healthcare providers Medical device makers
Industry sector
3345 Medical Device Manufacturing 6211 Healthcare Providers
Activity scope
Medical Device Usage Clinical Hemodynamic Monitoring Cardiac Support Device Management
Threshold
Impella RP with SmartAssist (Product Code 0046-0035) and Impella RP Flex with SmartAssist (Product Code 1000323)
Geographic scope
United States US

Taxonomy

Primary area
Medical Devices
Operational domain
Clinical Operations
Compliance frameworks
FDA 21 CFR Part 11
Topics
Healthcare Patient Safety

Get Pharma & Drug Safety alerts

Weekly digest. AI-summarized, no noise.

Free. Unsubscribe anytime.

Get alerts for this source

We'll email you when FDA Medical Device Recalls publishes new changes.

Optional. Personalizes your daily digest.

Free. Unsubscribe anytime.