Changeflow GovPing Pharma & Drug Safety Draeger Atlan A350 and A350 XL Anesthesia Works...
Urgent Enforcement Amended Final

Draeger Atlan A350 and A350 XL Anesthesia Workstation Correction

Favicon for www.fda.gov FDA Medical Device Recalls
Filed
Detected
Email

Summary

FDA classified Draeger Atlan A350 and A350 XL anesthesia workstations as a Class I recall—the most serious type—due to potential piston ventilator failure before or during use. Draeger is instructing customers to inspect and, if necessary, replace the ventilator motor assembly. Healthcare providers may continue using affected devices under permanent supervision until the corrective action is implemented.

Published by FDA on fda.gov . Detected, standardized, and enriched by GovPing. Review our methodology and editorial standards .

What changed

FDA classified this Draeger recall as Class I—the most serious type—after determining that piston ventilator failures in Atlan A350 and A350 XL anesthesia workstations may cause serious injury or death. Draeger traced the failures to impurities introduced during manufacturing and is instructing customers to inspect and replace the ventilator motor assembly. Customers may continue operating affected devices under permanent supervision until corrective action is completed, using manual ventilation if mechanical ventilation fails.

Healthcare providers using these anesthesia workstations must ensure all staff are aware of the correction notice and implement inspection/replacement procedures. The FDA emphasizes that continuing to use affected devices without correction poses serious risk of patient harm including hypoxia and cardiac arrest if mechanical ventilation fails during a procedure.

What to do next

  1. Inspect and replace the ventilator motor assembly on affected Atlan A350 and A350 XL devices
  2. Operate affected devices only under permanent supervision until corrective action is completed
  3. Return the Medical Device Correction Return Response Acknowledgment Form to Draeger

Archived snapshot

Apr 15, 2026

GovPing captured this document from the original source. If the source has since changed or been removed, this is the text as it existed at that time.


This recall **involves correcting certain devices and does not involve removing them from where they are used or sold.* The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without correction. This recall was classified on October 1, 2025.*

Affected Product

Product Name Model Number Unique Device Identifier
Atlan A350 8621500 04048675556176
Atlan A350 XL 8621600 04048675556183

Affected Serial Numbers

What to Do

Inspect and, if necessary, replace the ventilator motor assembly in all affected devices.

On September 9, 2025, Draeger, Inc. sent an Urgent Medical Device Correction to customers with instructions on how to correct the device if an error occurs:

  • If the error occurs before use, e.g. in standby or system test mode, mechanical ventilation cannot be started.
  • If the error occurs during a procedure, the mechanical ventilation may fail, and the device alerts the user by displaying the "Ventilator failure!!!" message. Manual ventilation or spontaneous breathing remains possible in both cases. It may be necessary to manually ventilate the patient to prevent patient injury. With the ongoing alarm of the ventilation failure, the Atlan continues to provide the Man/Spont ventilation mode, fresh-gas and agent delivery as well as all monitoring functions without restriction.
  • The corrective action is to inspect, and, if necessary, replace the ventilator motor assembly. You will be contacted by your local Draeger Service Representative to arrange a date for the ventilator motor assembly inspection and, if necessary, replacement.
  • Until the corrective action is implemented, you may continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use.
    • If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Monitor the patient’s condition; specifically, their oxygenation status as a brief cessation of ventilation can potentially lead to e.g., hypoxia, loss of lung recruitment, bradycardia and cardiac arrest.
  • If preferred, the alarm priority of the "Ventilator error!!!" can be downgraded with "ALARM RESET" after switching to Man/Spont ventilation mode.
  • Please ensure that all users of the Draeger Atlan as well as other persons concerned within your organization are made aware of this Urgent Medical Device Correction notice. Please complete and return the attached Medical Device Correction Return Response Acknowledgment and Receipt Form to confirm that you have received this information. If you have made the products available to third parties, please forward this Urgent Medical Device Correction notice to them.
  • Please retain this Urgent Medical Device Correction notice at least until your ventilator motor is inspected/replaced.

Reason for Correction

Draeger has received reports of the Atlan anesthesia workstation indicating a failure of the piston ventilator before use, and failures of the mechanical ventilation during use. After an investigation, Draeger was able to conclude that the errors were due to impurities introduced for a certain time frame during the manufacturing process.

As of September 9, 2025, Draeger has not reported any serious injuries or deaths related to this issue.

Device Use

The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates. The device can be used for mechanical ventilation, manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing. The device is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions

Contact Information

Customers in the U.S. with questions about this recall should contact Draeger Service Technical Support at 1-800-437-2437 (press two at the prompt, then two, then two again) between the hours of 8 am and 8 pm (EST).

Additional FDA Resources:

Unique Device Identifier (UDI)

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.

  • ## Content current as of:

04/14/2026

  • Regulated Product(s)

    • Medical Devices

Get daily alerts for FDA Medical Device Recalls

Daily digest delivered to your inbox.

Free. Unsubscribe anytime.

About this page

What is GovPing?

Every important government, regulator, and court update from around the world. One place. Real-time. Free. Our mission

What's from the agency?

Source document text, dates, docket IDs, and authority are extracted directly from FDA.

What's AI-generated?

The summary, classification, recommended actions, deadlines, and penalty information are AI-generated from the original text and may contain errors. Always verify against the source document.

Last updated

Classification

Agency
FDA
Filed
October 1st, 2025
Instrument
Enforcement
Legal weight
Binding
Stage
Final
Change scope
Substantive

Who this affects

Applies to
Medical device makers Healthcare providers
Industry sector
3345 Medical Device Manufacturing
Activity scope
Device correction response Medical device inspection
Geographic scope
United States US

Taxonomy

Primary area
Medical Devices
Operational domain
Clinical Operations
Compliance frameworks
GxP
Topics
Product Safety Healthcare

Get alerts for this source

We'll email you when FDA Medical Device Recalls publishes new changes.

Free. Unsubscribe anytime.

You're subscribed!