Class 1 Device Recall: Medline Medical Convenience Kits with Rotating Syringe Adaptor
Summary
FDA classified a Class 1 recall for Medline Industries medical convenience kits containing NAMIC Angiographic Rotating Adaptor control syringes. The firm identified through post-market surveillance a potential risk that the syringe rotating adaptor may unwind during use, resulting in loose connection or full disconnection between the syringe and manifold. Affected products include three SKUs (DYNJ57032D, DYNJ64026C, VASC1650) with 23 lot numbers totaling 2,154 units distributed US nationwide and internationally to 11 countries including Canada, Australia, Japan, and EU member states.
What changed
FDA classified a Class 1 (most serious) recall for Medline Industries medical convenience kits containing NAMIC Angiographic Rotating Adaptor control syringes. The recall was initiated after post-market surveillance revealed potential unwinding of the syringe rotating adaptor during use, which may cause loose connection or complete disconnection between the syringe and manifold, posing serious patient safety risks. The affected products span three SKUs with 23 different lot numbers, totaling 2,154 units distributed both domestically and internationally.
Healthcare facilities and distributors who received these kits must immediately quarantine affected inventory, complete the recall response form, and remove/discard the defective syringes before using any of the affected convenience kits. Medline will provide over-labels for remaining inventory and instructions for proper component removal. Distributors are required under FDA regulations to notify downstream customers of this recall. Affected parties should contact the Medline Recall Department at 866-359-1704 or recalls@medline.com to complete their response and receive remediation materials.
What to do next
- Immediately quarantine all affected product
- Complete the recall response form with affected product quantities
- Remove and discard affected Namic RA syringes from all kits prior to use
- Distributors must notify customers of this recall communication
Archived snapshot
Apr 16, 2026GovPing captured this document from the original source. If the source has since changed or been removed, this is the text as it existed at that time.
Class 1 Device Recall Medline medical convenience kits
- Databases
| 510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards |
| CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | |
| New Search | ) |
| | Class 1 Device Recall Medline medical convenience kits | |
| Date Initiated by Firm | February 27, 2026 |
|---|---|
| Date Posted | April 07, 2026 |
| Recall Status | Open , Classified |
| Recall Number | Z-1726-2026 |
| Recall Event ID | 98598 |
| Product Classification | Neurological tray - Product Code OJG |
| Product | Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: |
- ANGIO PACK, Medline SKU # DYNJ57032D;
- NEURO INTERVENTIONAL RADIOLOGY, Medline SKU # DYNJ64026C;
- NEURO INERVENTIONAL RADIOLOGY, Medline SKU # VASC1650. | | Code Information | Medline SKU # DYNJ57032D, UDI/DI each 10198459183850, UDI/DI case 40198459183851, Lot Number: 25HBN999; Medline SKU # DYNJ57032D, UDI/DI each 10198459183850, UDI/DI case 40198459183851, Lot Number: 25HBO039; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24JBC741; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24IBM270; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24IBE756; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24GBC735; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24FBF002; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24DBC238; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24ABY360; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 23LBH735; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26ABP305; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26ABI725; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25KBD893; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25HBM531; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25HBA055; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25EBC929; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBL096; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBD577; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBI493; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25ABI074; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 24LBG466; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26BBG430; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25LBC439. | | FEI Number | 1417592 | | Recalling Firm/ Manufacturer | Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753 | | For Additional Information Contact | Haley Barclay 866-359-1704 | | Manufacturer Reason for Recall | Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold. | | FDA Determined Cause | Under Investigation by firm | | Action | Medline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient at significant risk. Required Action: Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product.
If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 2154 units |
| Distribution | Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK). |
| Total Product Life Cycle | TPLC Device Report |
| 1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes. | |
Related changes
Get daily alerts for FDA Device Recalls (Class I)
Daily digest delivered to your inbox.
Free. Unsubscribe anytime.
Source
About this page
Every important government, regulator, and court update from around the world. One place. Real-time. Free. Our mission
Source document text, dates, docket IDs, and authority are extracted directly from FDA.
The summary, classification, recommended actions, deadlines, and penalty information are AI-generated from the original text and may contain errors. Always verify against the source document.
Classification
Who this affects
Taxonomy
Browse Categories
Get alerts for this source
We'll email you when FDA Device Recalls (Class I) publishes new changes.
Subscribed!
Optional. Filters your digest to exactly the updates that matter to you.