Changeflow GovPing Drug Safety Instinct Plus Endoscopic Clipping Device Recall...
Urgent Enforcement Removed Final

Instinct Plus Endoscopic Clipping Device Recalled Due to Malfunctions

Favicon for recalls-rappels.canada.ca Health Canada Recalls & Safety Alerts
Filed March 13th, 2026
Detected March 14th, 2026
Email

Summary

Health Canada has issued a Type II recall for the Instinct® Plus Endoscopic Clipping Device manufactured by Wilson-Cook Medical, Inc. due to malfunctions where the clip may remain attached to the drive wire and cannot be opened. This recall affects products manufactured prior to corrective actions being implemented.

What changed

Health Canada has issued a Type II recall for the Instinct® Plus Endoscopic Clipping Device due to a performance issue where the clip housing can detach from the catheter attachment, leaving the clip attached to the drive wire and unable to open. The manufacturer, Wilson-Cook Medical, Inc., has received 486 complaints regarding this malfunction. This recall specifically targets products manufactured before corrective actions were implemented to address this failure mode.

Healthcare providers in Canada are advised to contact the manufacturer, Wilson-Cook Medical, Inc., for additional information regarding this recall. While no specific compliance deadline is mentioned, the nature of a product recall implies that affected entities should cease using the non-compliant devices and consult with the manufacturer to understand the scope and implications of the recall. The recall class is Type II, indicating that use of or exposure to the faulty product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

What to do next

  1. Consult with Wilson-Cook Medical, Inc. regarding the Instinct® Plus Endoscopic Clipping Device recall.
  2. Cease use of affected Instinct® Plus Endoscopic Clipping Devices manufactured prior to corrective actions.
  3. Report any health or safety concerns related to this device to Health Canada.

Source document (simplified)

Health product recall

Instinct® Plus Endoscopic Clipping Device

Brand(s)

Wilson-Cook Medical, Inc.

Last updated

2026-03-13

Summary

Product Instinct® Plus Endoscopic Clipping Device Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information.

Audience Healthcare Industry

Affected products

| Affected products | Lot or serial number | Model or catalogue number |
| --- | --- | --- |
| Instinct® Plus Endoscopic Clipping Device | More than 10 numbers, contact manufacturer. | INSC-P-7-230-S |

Issue

Cook Medical has received 486 complaints of a malfunction that occurs when the user attempts to open the clip jaws by actuating the handle. The malfunction that may occur is the clip housing detaching from the catheter attachment and the clip remaining attached to the drive wire. The clip will extend from the catheter, but it will remain attached to the internal drive wire instead of opening the clip jaws. The clip cannot be opened if this occurs.

Corrective actions have been implemented to mitigate the risk of this failure mode.  This action is being taken to withdraw products manufactured prior to the implementation of the corrective actions.

Additional information

Details

Original published date:

2026-03-13

Alert / recall type Health product recall Category Health products - Medical devices - Cardiovascular Companies Wilson-Cook Medical, Inc.

4900 Bethania Station Rd., Winston Salem, North Carolina, United States, 27105

Published by Health Canada Audience Healthcare Industry Recall class Type II Recall date

2026-03-03

Identification number RA-81736

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe

Report a health or safety concern

Source

Analysis generated by AI. Source diff and links are from the original.

Classification

Agency
Various Canadian Agencies
Filed
March 13th, 2026
Instrument
Enforcement
Legal weight
Binding
Stage
Final
Change scope
Substantive

Who this affects

Applies to
Healthcare providers
Geographic scope
National (CA)

Taxonomy

Primary area
Healthcare
Operational domain
Compliance
Topics
Medical Devices Product Safety