Olympus Corporation of the Americas
Device Recalls (Class II)
Class II
Z-1478-2026
·
20260311
·
Ongoing
Product
Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy ...
Reason for Recall
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.
Source: openFDA Enforcement API